architecture

我们对两款光学追踪系统进行了对比,旨在评估它们在配合基准针进行软组织导航时的适用性:即采用无源标记的 Polaris 系统(由 Northern Digital Inc. (NDI) 制造;位于加拿大安大略省滑铁卢)以及 MicronTracker 2(型号 H40,由 Claron Technology, Inc. 制造;位于加拿大安大略省多伦多)。我们介绍了相应的工具设计方案,并在具有足够测量空间且处于典型临床照明环境的条件下,对工具尖端的追踪精度进行了评估。为了考察这些追踪系统的鲁棒性,我们进一步评估了它们对照明条件,以及被追踪工具的移动速度和空间姿态的敏感程度。结果显示,Polaris 系统在所有测试条件下均保持了稳健的追踪精度,而 MicronTracker 2 则对上述各项因素表现出高度的敏感性。




Comparative assessment of optical tracking systems for soft tissue navigation with fiducial needles
L. Maier-HeinA. FranzH.-P. MeinzerI. Wolf
Abstract
We compare two optical tracking systems with regard to their suitability for soft tissue navigation with fiducial needles: The Polaris system with passive markers (Northern Digital Inc. (NDI); Waterloo, Ontario, Canada), and the MicronTracker 2, model H40 (Claron Technology, Inc.; Toronto, Ontario, Canada). We introduce appropriate tool designs and assess the tool tip tracking accuracy under typical clinical light conditions in a sufficiently sized measurement volume. To assess the robustness of the tracking systems, we further evaluate their sensitivity to illumination conditions as well as to the velocity and the orientation of a tracked tool. While the Polaris system showed robust tracking accuracy under all conditions, the MicronTracker 2 was highly sensitive to the examined factors.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.